IFD2415-3
From the confocalDT IFD2415 series
Integrated confocal sensor and controller for distance and thickness measurement of up to five transparent layers. The measuring rate is adjustable up to 25 kHz.
- Number of measurement channels
- 1
Technical specification
Every engineering figure, laid out to scan quickly
| Measuring range | 3 mm |
|---|---|
| Start of measuring range | 20 mm |
| Linearity | 0.6 µm |
| Measuring rate | 25 kHz |
| Light spot diameter | 0.009 mm |
| Maximum measuring angle | 24 deg |
| Numerical aperture (NA) | 0.45 |
| Layers in multi-peak measurement | 5 layers |
| Power consumption | 7 W |
| Outputs and interfaces | Digital I/O |
| Number of measurement channels | 1 |
| Storage temperature | -20–70 °C |
| Operating temperature | 5–50 °C |
| Weight | 0.6 kg |
| Electrical connection | 12-pin M12 for supply, Ethernet, RS422 and sync; 17-pin M12 for I/O, analog and encoder; optional 3/6/9/15 m cable extension. |
| Mounting | Radial clamping (see accessories for mounting adapter), threaded holes |
| Permissible ambient light | 30,000 lx |
| Technical specification details | 3 mm net measuring range |
| Accuracy | 0.6 µm |
| Minimum target thickness | 0.15 mm |
| Target material | Reflective, diffuse as well as transparent surfaces (e.g. glass) |
| Sampling rate reference | IFD2415 integrated controller: up to 25 kHz; multi-peak thickness mode is documented for transparent layers |
| Linearity reference | https://www.micro-epsilon.com/fileadmin/download/products/dat--confocalDT-IFD2415--en.pdf published optical-probe linearity; axial resolution is not substituted for linearity. |
| Contact | Non-contact |
Values by configuration
These values differ with the controller or interface the sensor is used with. Each row applies to one configuration only; check the configuration in the system builder or with us.
| Interfaces | EtherNet/IP |
|---|---|
| Interfaces | EtherCAT |
| Interfaces | RS422 |
| Interfaces | Ethernet |
| Interfaces | PROFINET |
| Output resolution | 18 bit |
| Output resolution | 32 bit |
| Protection class (DIN EN 60529) | IP65 |
| Protection class (DIN EN 60529) | IP64 |
| Resolution | 0.015 µm |
| Resolution | 0.008 µm |
| Resolution | 0.08 µm |
Compatible accessories
Cables
Power supplies
Complementary products
The industries we serve
Frequently asked questions
Answers to the most common questions about our products and services
The measuring span is 3 mm, starting approximately 20 mm from the sensor face. Mount the sensor so target motion and height changes stay inside this window.
The datasheet specifies thickness measurement of up to five transparent layers. Practical suitability depends on resolving the interface returns and setting the materials’ refractive indices.
The rate is adjustable from 100 Hz to 25 kHz. For production-line integration, also verify the communication settings and the rate at which the receiving controller gets data.
Get a quote or advice
Tell us what you need and an engineer will come back to you.