TB2 Torque
Non-rotating flange-mount torque reference transducer with passive mV/V output. IP67 is available as a configuration option.
This series on the HBK site (opens in a new tab)
Shared technical specifications for the series
Properties every model in the series has. What differs is in the models table below.
Range across models
The span covered by the models of the series together — not the range of each model.
| Torque range | -10000–10000 Nm |
|---|---|
| ADC sampling rate | 450–40000 samples/s |
Models in the series
One column per model: what differs between the models of the series.
| Parameter | 1-TB2/100NM | 1-TB2/10KNM | 1-TB2/1KNM | 1-TB2/200NM | 1-TB2/2KNM | 1-TB2/3KNM | 1-TB2/500NM | 1-TB2/5KNM | TB2-100Nm | TB2-10kNm | TB2-1kNm | TB2-200Nm | TB2-2kNm | TB2-500Nm | TB2-5kNm |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Rotation | Non-rotating | Non-rotating | Non-rotating | Non-rotating | Non-rotating | Non-rotating | Non-rotating | Non-rotating | Non-rotating | Non-rotating | Non-rotating | Non-rotating | Non-rotating | Non-rotating | Non-rotating |
| Mechanical connection | Flange | Flange | Flange | Flange | Flange | Flange | Flange | Flange | Flange | Flange | Flange | Flange | Flange | Flange | Flange |
| Torque range | -100–100 Nm | -10000–10000 Nm | -1000–1000 Nm | -200–200 Nm | -2000–2000 Nm | -3000–3000 Nm | -500–500 Nm | -5000–5000 Nm | -100–100 Nm | -10000–10000 Nm | -1000–1000 Nm | -200–200 Nm | -2000–2000 Nm | -500–500 Nm | -5000–5000 Nm |
| Sampling rate reference | The selected HBK conditioner, filter and DAQ determine the digital sample rate. | The selected HBK conditioner, filter and DAQ determine the digital sample rate. | The selected HBK conditioner, filter and DAQ determine the digital sample rate. | The selected HBK conditioner, filter and DAQ determine the digital sample rate. | The selected HBK conditioner, filter and DAQ determine the digital sample rate. | The selected HBK conditioner, filter and DAQ determine the digital sample rate. | The selected HBK conditioner, filter and DAQ determine the digital sample rate. | The selected HBK conditioner, filter and DAQ determine the digital sample rate. | DMP41 selectable result rate, intended for reference and calibration measurements | MX840B maximum per-channel sample rate; sensor mechanics and filter settings can reduce usable bandwidth | MX840B maximum per-channel sample rate; sensor mechanics and filter settings can reduce usable bandwidth | ClipX fixed internal sampling rate; verify the configured filter and usable output bandwidth | ClipX fixed internal sampling rate; verify the configured filter and usable output bandwidth | DMP41 selectable result rate, intended for reference and calibration measurements | ClipX fixed internal sampling rate; verify the configured filter and usable output bandwidth |
| Linearity reference | No single accuracy class is inferred when it is not shown for the exact ordering code. | No single accuracy class is inferred when it is not shown for the exact ordering code. | No single accuracy class is inferred when it is not shown for the exact ordering code. | No single accuracy class is inferred when it is not shown for the exact ordering code. | No single accuracy class is inferred when it is not shown for the exact ordering code. | No single accuracy class is inferred when it is not shown for the exact ordering code. | No single accuracy class is inferred when it is not shown for the exact ordering code. | No single accuracy class is inferred when it is not shown for the exact ordering code. | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified |
| Technical specification details | Actions: Not sellable online; Measurement range: 885 in-lb; Size (length): 3.31 in | Actions: Not sellable online; Measurement range: 88507 in-lb; Size (length): 7.72 in | Actions: Not sellable online; Measurement range: 8851 in-lb; Size (length): 4 in | Actions: Not sellable online; Measurement range: 1770 in-lb; Size (length): 3.31 in | Actions: Not sellable online; Measurement range: 17701 in-lb; Size (length): 5.12 in | Actions: Not sellable online; Measurement range: 26552 in-lb; Size (length): 5.12 in | Actions: Not sellable online; Measurement range: 4425 in-lb; Size (length): 4 in | Actions: Not sellable online; Measurement range: 44254 in-lb; Size (length): 6.12 in | ±100 Nm rated torque; static / reaction installation | ±10 kNm rated torque; static / reaction installation | ±1 kNm rated torque; static / reaction installation | ±200 Nm rated torque; static / reaction installation | ±2 kNm rated torque; static / reaction installation | ±500 Nm rated torque; static / reaction installation | ±5 kNm rated torque; static / reaction installation |
| Outputs and interfaces | Unknown | Unknown | Unknown | Unknown | Unknown | Unknown | Unknown | Unknown | Digital I/O | RS422 / RS485 | Analog 0-10V | RS422 / RS485 | RS422 / RS485 | RS422 / RS485 | RS422 / RS485 |
| Rotational speed measurement | False | False | False | False | False | False | False | False | False | False | False | False | False | False | False |
| Repeatability reference | Use the individual calibration certificate and installed-chain verification; no unsupported scalar repeatability is inferred. | Use the individual calibration certificate and installed-chain verification; no unsupported scalar repeatability is inferred. | Use the individual calibration certificate and installed-chain verification; no unsupported scalar repeatability is inferred. | Use the individual calibration certificate and installed-chain verification; no unsupported scalar repeatability is inferred. | Use the individual calibration certificate and installed-chain verification; no unsupported scalar repeatability is inferred. | Use the individual calibration certificate and installed-chain verification; no unsupported scalar repeatability is inferred. | Use the individual calibration certificate and installed-chain verification; no unsupported scalar repeatability is inferred. | Use the individual calibration certificate and installed-chain verification; no unsupported scalar repeatability is inferred. | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified |
| Sampling rate limited by system readout | False | False | False | False | False | False | False | False | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified |
| Accuracy class | 0.02% | 0.02% | 0.02% | 0.02% | 0.02% | 0.02% | 0.02% | 0.02% | 0.02% | 0.02% | 0.02% | 0.02% | 0.02% | 0.02% | 0.02% |
| Numeric manufacturer accuracy class | 0.02 | 0.02 | 0.02 | 0.02 | 0.02 | 0.02 | 0.02 | 0.02 | 0.02 | 0.02 | 0.02 | 0.02 | 0.02 | 0.02 | 0.02 |
| Interfaces | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified | Ethernet | EtherCAT | EtherCAT | EtherCAT | EtherCAT | Ethernet | Ethernet |
| Accuracy (% FSO) | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified | 0.02 % | 0.02 % | 0.02 % | 0.02 % | 0.02 % | 0.02 % | 0.02 % |
| ADC sampling rate | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified | Not specified | 19200 samples/s | 450 samples/s | 40000 samples/s | 40000 samples/s | 40000 samples/s | 40000 samples/s | 450 samples/s |
| Status | Active | Active | Active | Active | Active | Active | Active | Active | Active | Active | Active | Active | Active | Active | Active |
The industries we serve
Frequently asked questions
Answers to the most common questions about our products and services
No. The table specifies standard IP54 and optional IP67. Associate the rating with the configuration rather than the series name alone.
Accuracy class 0.02 differs from classification under DIN 51309 or EA-10/14. The latter is documented in conjunction with an appropriate calibration certificate.
No. The table separates torque intervals of 0–20%, above 20–60%, and above 60–100% of nominal torque.
The table specifies 1 mV/V nominal output and bridge excitation. Use a compatible amplifier and preserve the particular transducer’s sensitivity and calibration.
No. The nominal range is +10 to +60 °C, while the operating range is −10 to +80 °C. For reference measurements, check temperature effects on zero and sensitivity as well as operating limits.
Get a quote or advice
Tell us what you need and an engineer will come back to you.