TYPE 4326

Type 4326 measures acceleration along three axes with individually calibrated outputs and a nominal charge sensitivity of 3.10 pC/g per axis. The 4326-A is lighter, while the 4326-A-001 is rated for higher operating temperatures. Mounting depends on the surface and the required frequency band.

Visit the HBK site (opens in a new tab)

Shared technical specifications for the series

Properties every model in the series has. What differs is in the models table below.

Range across models

The span covered by the models of the series together — not the range of each model.

Range across the models of TYPE 4326
Frequency response 0.64–16000000 Hz
Operating temperature -60–288 °C

Models in the series

One column per model: what differs between the models of the series.

Models of TYPE 4326 compared by parameter
Parameter -4326–001- -4326--001- -4326-A– -4326-A-- -4326-A-001-
Technical specification details ±3000 g indexed capacity; Triaxial; Miniature/ESS; Clip mounting ±3000 g indexed capacity; Triaxial; General purpose; Clip mounting ±3000 g indexed capacity; Triaxial; General purpose; Clip mounting
Frequency response 0.64–16000000 Hz 0.64–10000 Hz 0.64–10000 Hz
Operating temperature -60–288 °C -60–175 °C -60–288 °C
Vibration sensor output Piezoelectric charge; Analogue Piezoelectric charge; Analogue Piezoelectric charge; Analogue
Sampling rate reference Sensor bandwidth is indexed by HBK; the selected conditioner and DAQ determine digital sample rate. Sensor bandwidth is indexed by HBK; the selected conditioner and DAQ determine digital sample rate. Sensor bandwidth is indexed by HBK; the selected conditioner and DAQ determine digital sample rate.
Static (DC) acceleration measurement False False False
Repeatability reference HBK does not publish a separate scalar repeatability for every ordering code; use individual calibration and controlled mounting. HBK does not publish a separate scalar repeatability for every ordering code; use individual calibration and controlled mounting. HBK does not publish a separate scalar repeatability for every ordering code; use individual calibration and controlled mounting.
Outputs and interfaces Unknown Unknown Unknown
Acceleration sensitivity 3.1 mV/g 3.1 mV/g 3.1 mV/g
Sampling rate limited by system readout True True True
Linearity reference Official indexed sensitivity tolerance: 20%. Official indexed sensitivity tolerance: 20%. Official indexed sensitivity tolerance: 20%.
Number of axes 3 3 3
Number of measurement channels 3 3 3
Status Active Active Active

The industries we serve

Frequently asked questions

Answers to the most common questions about our products and services

The titanium 4326-A-001 weighs 17 g and is rated up to 230°C. The aluminum 4326-A weighs 13 g and is rated up to 175°C.

The datasheet specifies 1 Hz to 9 kHz on X, to 8 kHz on Y and to 16 kHz on Z at the ±10% amplitude limit. The charge amplifier sets the practical low-frequency limit.

HBK describes adhesive, clip, M2 screw and M3 stud mounting. Choose based on surface geometry and the required high-frequency response.

The sensing elements are electrically isolated from the common housing. This is useful in multichannel measurements where differences in ground potential can add noise.

Connect each 10–32 UNF output to a dedicated charge channel or compatible charge converter. Keep the axis label, acquisition channel and calibration factor matched.

Didn't find the answer you were looking for?

Talk to an expert

More information

Get a quote or advice

Tell us what you need and an engineer will come back to you.

Request a quote for this product
What would you like to ask for?

Email or phone — one of them is enough

Email or phone — one of them is enough

0 of 500 words