IFS2406-10
From the IFS2406 series
Confocal heads for distance and thickness measurement, including dedicated vacuum versions. Beam geometry depends on the model.
- Number of measurement channels
- 2
Technical specification
Every engineering figure, laid out to scan quickly
| Measuring range | 10 mm |
|---|---|
| Start of measuring range | 27 mm |
| Linearity | 2 µm |
| Light spot diameter | 0.015 mm |
| Maximum measuring angle | 13.5 deg |
| Numerical aperture (NA) | 0.25 |
| Storage temperature | -20–70 °C |
| Operating temperature | 5–70 °C |
| Weight | 0.128 kg |
| Mounting | Radial clamping (mounting adapter see accessories) |
| Technical specification details | 10 mm net measuring range |
| Accuracy | 2.4 µm |
| Minimum target thickness | 0.5 mm |
| Target material | reflective, diffuse as well as transparent surfaces (e.g. glass) |
| Linearity reference | https://www.micro-epsilon.com/fileadmin/download/products/cat--confocalDT--en-us.pdf published optical-probe linearity; axial resolution is not substituted for linearity. |
| Contact | Non-contact |
Values by configuration
These values differ with the controller or interface the sensor is used with. Each row applies to one configuration only; check the configuration in the system builder or with us.
| Interfaces | Ethernet |
|---|---|
| Interfaces | EtherCAT |
| Interfaces | PROFINET |
| Number of measurement channels | 2 |
| Number of measurement channels | 1 |
| Outputs and interfaces | Analog 0-10V |
| Outputs and interfaces | RS422 / RS485 |
| Outputs and interfaces | Digital I/O |
| Outputs and interfaces | 4-20mA |
| Protection class (DIN EN 60529) | IP40 |
| Protection class (DIN EN 60529) | IP64 |
| Protection class (DIN EN 60529) | IP65 |
| Resolution | 0.019 µm |
| Resolution | 0.207 µm |
| Sampling rate reference | IFC2422MP: 10,000 measurements/s per synchronous channel |
| Sampling rate reference | IFC2421: 10,000 measurements/s |
| Sampling rate reference | IFC2466: 30,000 measurements/s per synchronous channel |
| Sampling rate reference | IFC2417: 25,000 measurements/s per synchronous channel |
| Sampling rate reference | IFC2411: 8,000 measurements/s |
| Sampling rate reference | IFC2412: 8,000 measurements/s per synchronous channel |
| Sampling rate reference | IFC2421MP: 10,000 measurements/s |
| Sampling rate reference | IFC2422: 10,000 measurements/s per synchronous channel |
| Sampling rate reference | IFC2465: 30,000 measurements/s |
| Sampling rate reference | IFC2465MP: 30,000 measurements/s |
| Sampling rate reference | IFC2466MP: 30,000 measurements/s per synchronous channel |
| Sampling rate reference | IFC2416: 25,000 measurements/s |
Compatible system components
Controllers
-
IFC2411
Required
- Measuring rate
- 8 kHz
- Communication
- Ethernet, RS422
- Inputs
- 1
- Signal type
- Analog voltage, Analog current, RS422, Digital (EtherCAT/Ethernet)
-
IFC2411/IE
Required
- Measuring rate
- 8 kHz
- Communication
- RS422, PROFINET, EtherNet/IP, EtherCAT
- Inputs
- 1
- Signal type
- Analog voltage, Analog current, RS422, Digital (EtherCAT/Ethernet)
-
IFC2412
Required
- Measuring rate
- 8 kHz
- Communication
- RS422, Ethernet
- Inputs
- 2
- Signal type
- Analog voltage, Analog current, RS422, Digital (EtherCAT/Ethernet)
-
IFC2412/IE
Required
- Measuring rate
- 8 kHz
- Communication
- EtherCAT, RS422
- Inputs
- 2
- Signal type
- Analog voltage, Analog current, RS422, Digital (EtherCAT/Ethernet)
-
IFC2416
Required
- Measuring rate
- 25 kHz
- Communication
- Ethernet, RS422
- Inputs
- 1
- Signal type
- Analog voltage, Analog current, RS422, Digital (EtherCAT/Ethernet)
-
IFC2417
Required
- Measuring rate
- 25 kHz
- Communication
- RS422, Ethernet
- Inputs
- 2
- Signal type
- Analog voltage, Analog current, RS422, Digital (EtherCAT/Ethernet)
-
IFC2417/IE
Required
- Measuring rate
- 25 kHz
- Communication
- EtherCAT, RS422
- Inputs
- 2
- Signal type
- Analog voltage, Analog current, RS422, Digital (EtherCAT/Ethernet)
-
IFC2421
Required
- Measuring rate
- 10 kHz
- Communication
- Ethernet, RS422, EtherCAT
- Inputs
- 1
- Signal type
- Analog voltage, Analog current, RS422, Digital (EtherCAT/Ethernet)
-
IFC2421MP
Required
- Measuring rate
- 10 kHz
- Communication
- EtherCAT, Ethernet, RS422
- Inputs
- 1
- Signal type
- Analog voltage, Analog current, RS422, Digital (EtherCAT/Ethernet)
-
IFC2422
Required
- Measuring rate
- 10 kHz
- Communication
- EtherCAT, Ethernet, RS422
- Inputs
- 2
- Signal type
- Analog voltage, Analog current, RS422, Digital (EtherCAT/Ethernet)
-
IFC2422MP
Required
- Measuring rate
- 10 kHz
- Communication
- Ethernet, EtherCAT, RS422
- Inputs
- 2
- Signal type
- Analog voltage, Analog current, RS422, Digital (EtherCAT/Ethernet)
-
IFC2465
Required
- Measuring rate
- 30 kHz
- Communication
- Ethernet, EtherCAT, RS422
- Inputs
- 1
- Signal type
- Analog voltage, Analog current, RS422, Digital (EtherCAT/Ethernet)
-
IFC2465MP
Required
- Measuring rate
- 30 kHz
- Communication
- EtherCAT, RS422, Ethernet
- Inputs
- 1
- Signal type
- Analog voltage, Analog current, RS422, Digital (EtherCAT/Ethernet)
-
IFC2466
Required
- Measuring rate
- 30 kHz
- Communication
- EtherCAT, Ethernet, RS422
- Inputs
- 2
- Signal type
- Analog voltage, Analog current, RS422, Digital (EtherCAT/Ethernet)
-
IFC2466MP
Required
- Measuring rate
- 30 kHz
- Communication
- RS422, Ethernet, EtherCAT
- Inputs
- 2
- Signal type
- Analog voltage, Analog current, RS422, Digital (EtherCAT/Ethernet)
Compatible accessories
Optical fibres
-
C2401-3
Required with the chosen components
Standard optical fiber; length 3 m; FC/APC and E2000/APC connectors. For IFS2406-10.
-
C2401/PT3-3
Required with the chosen components
Optical fiber with protective hose for mechanical stress; length 3 m; FC/APC and E2000/APC connectors. For IFS2406-10.
-
C2401-3(10)
Required with the chosen components
Drag-chain compatible optical fiber; length 3 m; FC/APC and E2000/APC connectors. For IFS2406-10.
-
C2401-3(20)
Required with the chosen components
Robot-compatible optical fiber; length 3 m; FC/APC and E2000/APC connectors. For IFS2406-10.
-
C2400-x
Required with the chosen components
-
C2400/PT-x
Required with the chosen components
-
C2400/PT-x-Vac
Required with the chosen components
-
C2401-5
Required with the chosen components
Standard optical fiber; length 5 m; FC/APC and E2000/APC connectors. For IFS2406-10.
-
C2401-10
Required with the chosen components
Standard optical fiber; length 10 m; FC/APC and E2000/APC connectors. For IFS2406-10.
-
C2401/PT3-5
Required with the chosen components
Optical fiber with protective hose for mechanical stress; length 5 m; FC/APC and E2000/APC connectors. For IFS2406-10.
-
C2401/PT3-10
Required with the chosen components
Optical fiber with protective hose for mechanical stress; length 10 m; FC/APC and E2000/APC connectors. For IFS2406-10.
-
C2401-5(10)
Required with the chosen components
Drag-chain compatible optical fiber; length 5 m; FC/APC and E2000/APC connectors. For IFS2406-10.
-
C2401-10(10)
Required with the chosen components
Drag-chain compatible optical fiber; length 10 m; FC/APC and E2000/APC connectors. For IFS2406-10.
-
C2401-5(20)
Required with the chosen components
Robot-compatible optical fiber; length 5 m; FC/APC and E2000/APC connectors. For IFS2406-10.
-
C2401-10(20)
Required with the chosen components
Robot-compatible optical fiber; length 10 m; FC/APC and E2000/APC connectors. For IFS2406-10.
Power supplies
-
PS2020 (2420062)
Optional
Mounting
-
MA2400-27
Optional
-
JMA-27
Optional
-
JMA-Thickness-27
Optional
Accessories
-
C2402/Vac/KF16
Optional
-
C2405/Vac/1/KF16
Optional
-
C2405/Vac/1/CF16
Optional
-
C2405/Vac/6/CF63
Optional
Complementary products
-
IF1032/ETH
Optional
-
IF2001/USB
Optional
-
IF2004/USB
Optional
-
IF2008/PCIE
Optional
-
IF2008/ETH
Optional
-
IFL2422/LED
Optional
- IFL24x1/LED configurable component family Optional Only for source configuration configuration-a255bca6fcc79734c5f2548e: IFS2406-10; selected configuration must be matched before display.
Cables
-
SC2415-x/IF2008
Optional
-
SC2415-x/IF2008ETH
Optional
-
SC2415-x/OE
Optional
-
SC2471-x/IF2008
Optional
-
SC2471-3/IF2008ETH
Optional
-
SC2471-x/RS422/OE
Optional
The industries we serve
Frequently asked questions
Answers to the most common questions about our products and services
The measuring span is 10 mm and starts approximately 27 mm from the sensor face. Position the target so that the expected motion stays within the window.
The datasheet specifies less than 207 nm RMS noise at mid-span at a 1 kHz measuring rate. This value describes noise under the specified test conditions.
The datasheet lists 0.5 mm for glass with a refractive index of 1.5 under the specified test conditions. For a different material, verify that the controller can resolve the returns from both surfaces.