Clause 7.1.5 is very short, yet it is the most common audit finding in Israeli industry.

Amit Ben-David

מהנדס יישומים

Published Updated

ISO 9001:2015 gives monitoring and measuring resources less than a page. That brevity is exactly what creates the gaps: the standard requires an outcome, not a procedure, and leaves it to you to prove you achieved it.

What the standard actually requires

  • That equipment is calibrated or verified against a standard with international traceability, at defined intervals.
  • That calibration status is identified — meaning you can look at an instrument and know whether it is valid.
  • That the instrument is protected from adjustments that would invalidate its results.
  • That if an instrument is found out of tolerance, the validity of its previous measurements is reviewed and action taken.

The clause people fail on

The fourth. An instrument that comes back from calibration out of tolerance means every measurement made with it since the previous calibration is suspect. Organisations document the calibration and don’t document the retrospective review, and that is a finding — not because something bad happened, but because there is no way to show that it didn’t.

Where this touches a sensor on the line

A sensor used to release product is measuring equipment in every sense, even when it is built into a machine and nobody reads it by hand. If the machine rejects parts based on its reading, it is inside the system.

The first question an auditor asks isn’t “have you calibrated?”. It’s “show me the traceability certificate” — followed by “what did you do the time an instrument came back out of tolerance?”.